作者
A Uneri, T De Silva, J Goerres, MW Jacobson, MD Ketcha, S Reaungamornrat, G Kleinszig, S Vogt, AJ Khanna, GM Osgood, JP Wolinsky, JH Siewerdsen
发表日期
2017/3/28
期刊
Physics in Medicine & Biology
卷号
62
期号
8
页码范围
3330
出版商
IOP Publishing
简介
Intraoperative x-ray radiography/fluoroscopy is commonly used to assess the placement of surgical devices in the operating room (eg spine pedicle screws), but qualitative interpretation can fail to reliably detect suboptimal delivery and/or breach of adjacent critical structures. We present a 3D–2D image registration method wherein intraoperative radiographs are leveraged in combination with prior knowledge of the patient and surgical components for quantitative assessment of device placement and more rigorous quality assurance (QA) of the surgical product. The algorithm is based on known-component registration (KC-Reg) in which patient-specific preoperative CT and parametric component models are used. The registration performs optimization of gradient similarity, removes the need for offline geometric calibration of the C-arm, and simultaneously solves for multiple component bodies, thereby allowing QA …