作者
Shougang Wang, Wei-Ning Lee, Jean Provost, Jianwen Luo, Elisa E Konofagou
发表日期
2008/10
期刊
IEEE transactions on ultrasonics, ferroelectrics, and frequency control
卷号
55
期号
10
页码范围
2221-2233
出版商
IEEE
简介
High frame-rate ultrasound RF data acquisition has been proved to be critical for novel cardiovascular imaging techniques, such as high-precision myocardial elastography, pulse wave imaging (PWI), and electromechanical wave imaging (EWI). To overcome the frame-rate limitations on standard clinical ultrasound systems, we developed an automated method for multi-sector ultrasound imaging through retrospective electrocardiogram (ECG) gating on a clinically used open architecture system. The method achieved both high spatial (64 beam density) and high temporal resolution (frame rate of 481 Hz) at an imaging depth up to 11 cm and a 100% field of view in a single breath-hold duration. Full-view imaging of the left ventricle and the abdominal aorta of healthy human subjects was performed using the proposed technique in vivo. ECG and ultrasound RF signals were simultaneously acquired on a personal …
引用总数
2008200920102011201220132014201520162017201820192020202120222023202456101297137377643421
学术搜索中的文章
S Wang, WN Lee, J Provost, J Luo, EE Konofagou - IEEE transactions on ultrasonics, ferroelectrics, and …, 2008