Valerio Ricciardi, Pavel Bláha, Raffaele Buompane, Giuseppina Crescente, Giacomo Cuttone, Lucio Gialanella, Katarina Michaličková, Severina Pacifico, Giuseppe Porzio, Lorenzo Manti
发表日期
2021/1
期刊
Applied Sciences
卷号
11
期号
24
页码范围
11986
出版商
Multidisciplinary Digital Publishing Institute
简介
Featured Application
The application of the work described herein is twofold: the possibility of routinely irradiating biological samples with high accuracy in energy and dose at a low-energy particle accelerator to gain insights into fundamental mechanisms of the biological action of charged particles; in a broader scenario, the possibility to potentiate the therapeutic capabilities of protontherapy through a method based on a nuclear physics reaction.
Abstract
Protontherapy (PT) is a fast-growing cancer therapy modality thanks to much-improved normal tissue sparing granted by the charged particles’ inverted dose-depth profile. Protons, however, exhibit a low biological effectiveness at clinically relevant energies. To enhance PT efficacy and counteract cancer radioresistance, Proton–Boron Capture Therapy (PBCT) was recently proposed. PBCT exploits the highly DNA-damaging α-particles generated by the p + 11B→3α (pB) nuclear reaction, whose cross-section peaks for proton energies of 675 keV. Although a significant enhancement of proton biological effectiveness by PBCT has been demonstrated for high-energy proton beams, validation of the PBCT rationale using monochromatic proton beams having energy close to the reaction cross-section maximum is still lacking. To this end, we implemented a novel setup for radiobiology experiments at a 3-MV tandem accelerator; using a scattering chamber equipped with an Au foil scatterer for beam diffusion on the biological sample, uniformity in energy and fluence with uncertainties of 2% and 5%, respectively, was achieved. Human cancer cells were …
引用总数