Mustapha Bouhrara, David A Reiter, Hasan Celik, Kenneth W Fishbein, Richard Kijowski, Richard G Spencer
发表日期
2016/6
期刊
Magnetic resonance in medicine
卷号
75
期号
6
页码范围
2406-2420
简介
Purpose
To compare the reliability and stability of the multicomponent‐driven equilibrium single pulse observation of T1 and T2 (mcDESPOT) and Carl‐Purcell‐Meiboom‐Gill (CPMG) approaches to parameter estimation.
Methods
The stability and reliability of mcDESPOT and CPMG‐derived parameter estimates were compared through examination of energy surfaces, evaluation of model sloppiness, and Monte Carlo simulations. Comparisons were performed on an equal time basis and assuming a two‐component system. Parameter estimation bias, reflecting accuracy, and dispersion, reflecting precision, were derived for a range of signal‐to‐noise ratios (SNRs) and relaxation parameters.
Results
The energy surfaces for parameters incorporated into the mcDESPOT signal model exhibit flatness, a complex structure of local minima, and instability to noise to a much greater extent than the corresponding …