作者
Aaron Conway, Elizabeth Tipton, Wei-Hong Liu, Zachary Conway, Kathleen Soalheira, Joanna Sutherland, James Fingleton
发表日期
2019/2/1
来源
Thorax
卷号
74
期号
2
页码范围
157-163
出版商
BMJ Publishing Group Ltd
简介
Background
Transcutaneous carbon dioxide (TcCO2) monitoring is a non-invasive alternative to arterial blood sampling. The aim of this review was to determine the accuracy and precision of TcCO2 measurements.
Methods
Medline and EMBASE (2000–2016) were searched for studies that reported on a measurement of PaCO2 that coincided with a measurement of TcCO2. Study selection and quality assessment (using the revised Quality Assessment of Diagnostic Accuracy Studies tool (QUADAS-2)) were performed independently. The Grading Quality of Evidence and Strength of Recommendation approach was used to summarise the strength of the body of evidence. Pooled estimates of the mean bias between TcCO2 and PaCO2 and limits of agreement with outer 95% CIs (termed population limits of agreement) were calculated.
Results
The mean bias was −0.1 mm Hg and the population limits of agreement …