简介
A further study is conducted on two factors which respectively influence the sensitivity of optically pumped cesium magnetometer (CsOPM). The influence of radio frequency (RF) power and the buffer gas pressure on the sensitivity is theoretically analyzed, and some properties are predicted. Based on the established measurement system and the visible Zeeman spectrum, not only is the real influence of these factors studied, but also, under our experimental condition, optimum parameters based on the measured curves are ascertained. The properties of these measured curves match the theoretical result very well. Our research attempts to provide theory reference to help magnetometer designers determine optimum parameters under certain conditions.