作者
Stefan Gradl, Patrick Kugler, Clemens Lohmüller, Bjoern Eskofier
发表日期
2012/8/28
研讨会论文
2012 annual international conference of the IEEE engineering in medicine and biology society
页码范围
2452-2455
出版商
IEEE
简介
We developed an application for Android™-based mobile devices that allows real-time electrocardiogram (ECG) monitoring and automated arrhythmia detection by analyzing ECG parameters. ECG data provided by pre-recorded files or acquired live by accessing a Shimmer™ sensor node via Bluetooth™ can be processed and evaluated. The application is based on the Pan-Tompkins algorithm for QRS-detection and contains further algorithm blocks to detect abnormal heartbeats. The algorithm was validated using the MIT-BIH Arrhythmia and MIT-BIH Supraventricular Arrhythmia databases. More than 99% of all QRS complexes were detected correctly by the algorithm. Overall sensitivity for abnormal beat detection was 89.5% with a specificity of 80.6%. The application is available for download and may be used for real-time ECG-monitoring on mobile devices.
引用总数
20112012201320142015201620172018201920202021202220232024121024252627341819141094
学术搜索中的文章
S Gradl, P Kugler, C Lohmüller, B Eskofier - 2012 annual international conference of the IEEE …, 2012