作者
Yeh-Liang Hsu, Shih-Tseng Lee, Hao-Wei Lin
发表日期
2001/8/25
期刊
Biomedical Engineering: Applications, Basis and Communications
卷号
13
期号
04
页码范围
168-174
出版商
National Taiwan University
简介
Drilling tools currently used in surgery depends only on the surgeon's manual skills to stop the penetration when completing a hole. This paper presents a modular mechatronic system for automatic bone drilling in surgery. The development of a “modular system” that is compatible with motor-driven drills that are commercially available, rather than developing a new surgical drill, is emphasized. A fuzzy controller analyzes the electric current consumed by the DC motor of the drill. When break-through is detected, the power will be cut and stops drilling in order to prevent excessive protrusion of the drill bit. In extensive drilling tests on real human skulls, there were no unexpected failure, and the overshoots of all tests were well less than 2mm.
引用总数
200720082009201020112012201320142015201620172018201920202021202220232024111732564563221
学术搜索中的文章
YL Hsu, ST Lee, HW Lin - Biomedical Engineering: Applications, Basis and …, 2001