von Willebrand disease (VWD) is caused by a quantitative and/or qualitative deficiency of the von Willebrand factor (VWF). The laboratory diagnosis of VWD is dependent on the measurement of VWF antigen (VWF:Ag) and ristocetin cofactor activity (VWF:RCo). The aim of this study was to undertake a two‐centre evaluation of two new automated VWF:Ag and VWF:RCo assays systems from Instrumentation Laboratory (Bedford, USA). Using the two new analytical systems that operated with different detection principles: immunoturbidimetric (TOP500 analyser) and chemiluminescent (AcuStar analyser), VWF:Ag and VWF:RCo levels were determined in samples from 171 healthy normal subjects, 80 VWD patients (16 type 1, 58 type 2 and 6 type 3) and 7 acquired von Willebrand syndrome patients. With commercial lyophilized normal and pathological plasmas VWF: Ag and VWF:RCo assays performed on both …