T Varghese, J Ophir, I Cespedes
发表日期
1996/1/1
期刊
Ultrasound in medicine & biology
卷号
22
期号
8
页码范围
1043-1052
出版商
Elsevier
简介
Elastography uses estimates of the time delay (obtained by cross-correlation) to compute strain estimates in tissue due to quasistatic compression. Because the time delay estimates do not generally occur at the sampling intervals, the location of the cross-correlation peak does not give and accurate estimate of the time delay. Sampling errors in the time-delay estimate are reduced using signal interpolation techniques to obtain subsample time-delay estimates. Distortions of the echo signals due to tissue compression introduce correlation artifacts in the elastogram. These artifacts are reduced by a combination of small compressions and temporal stretching of the postcompression signal. Random noise effects in the resulting elastograms are reduced by averaging several elastograms, obtained from successive small compressions (assuming that the errors are uncorrelated). Multicompression averaging with …