作者
Sang Chul Lee, Ho Kyung Kim, In Kon Chun, Myung Hye Cho, Soo Yeol Lee, Min Hyoung Cho
发表日期
2003/12/1
期刊
Physics in Medicine & Biology
卷号
48
期号
24
页码范围
4173
出版商
IOP Publishing
简介
A dedicated small-animal x-ray micro computed tomography (micro-CT) system has been developed to screen laboratory small animals such as mice and rats. The micro-CT system consists of an indirect-detection flat-panel x-ray detector with a field-of-view of 120× 120 mm 2, a microfocus x-ray source, a rotational subject holder and a parallel data processing system. The flat-panel detector is based on a matrix-addressed photodiode array fabricated by a CMOS (complementary metal-oxide semiconductor) process coupled to a CsI: Tl (thallium-doped caesium iodide) scintillator as an x-ray-to-light converter. Principal imaging performances of the micro-CT system have been evaluated in terms of image uniformity, voxel noise and spatial resolution. It has been found that the image non-uniformity mainly comes from the structural non-uniform sensitivity pattern of the flat-panel detector and the voxel noise is about 48 …