作者
Aseem Kumar Tiwari, Prashant Pandey, Surbhi Dixit, Vimarsh Raina
发表日期
2012/3/1
期刊
Clinical chemistry and laboratory medicine
卷号
50
期号
3
页码范围
471-474
出版商
Walter de Gruyter
简介
Background: The objective of the study was to find the incidence of hemolysis in samples transported through a pneumatic tube system (PTS) at different speeds.
Methods: This prospective observational study was done in three phases: “short distance and high speed (115 m at 3 m/s)”, “long distance and high speed (225 m at 3 m/s)” and “short distance and slow speed (115 at 2 m/s)”. Fifty-two, 215 and 45 serum tube pairs, respectively, were evaluated in these three phases. A set of tubes was sent by PTS while the other was hand-carried. Samples were analyzed for supernatant hemoglobin (Hb), potassium (K+) and lactate dehydrogenase (LD).
Results: Mean transit time of samples through the PTS was much shorter as compared to human courier in all three phases. LD was elevated in PTS arm in the “short distance and high speed” phase and in the “long distance and high speed” phase, all three indices of …
引用总数
20122013201420152016201720182019202020212022202320242224184434591