作者
Ehsan Abadi, Brian Harrawood, Shobhit Sharma, Anuj Kapadia, William P Segars, Ehsan Samei
发表日期
2018/12/12
期刊
IEEE transactions on medical imaging
卷号
38
期号
6
页码范围
1457-1465
出版商
IEEE
简介
The purpose of this study was to develop a CT simulation platform that is: 1) compatible with voxel-based computational phantoms; 2) capable of modeling the geometry and physics of commercial CT scanners; and 3) computationally efficient. Such a simulation platform is designed to enable the virtual evaluation and optimization of CT protocols and parameters for achieving a targeted image quality while reducing radiation dose. Given a voxelized computational phantom and a parameter file describing the desired scanner and protocol, the developed platform DukeSim calculates projection images using a combination of ray-tracing and Monte Carlo techniques. DukeSim includes detailed models for the detector quantum efficiency, quantum and electronic noise, detector crosstalk, subsampling of the detector and focal spot areas, focal spot wobbling, and the bowtie filter. DukeSim was accelerated using GPU …
引用总数
20192020202120222023202471321221618
学术搜索中的文章