作者
Marvin M Doyley, John B Weaver, Elijah EW Van Houten, Francis E Kennedy, Keith D Paulsen
发表日期
2003/4
期刊
Medical Physics
卷号
30
期号
4
页码范围
495-504
出版商
American Association of Physicists in Medicine
简介
An objective contrast‐detail analysis was performed in this study to assess the low contrast detectability of a clinical prototype harmonic magnetic resonance elastographic imaging system. Elastographic imaging was performed on gelatin phantoms containing spherical inclusions of varying size and modulus contrast. The results demonstrate that lesions as small as 5 mm can be detected with a minimum modulus contrast of 14 dB. However, the shear modulus of such small lesions was not accurately recovered. In general, the shear modulus of larger focal lesions was accurately (i.e., within 25% of the true value) recovered. The minimum modulus contrast needed to detect focal lesions was observed to decrease with increasing lesion size.
引用总数