Stefan Pleus, Christina Schmid, ME Manuela Link, Eva Zschornack, Hans-Martin Klötzer, Cornelia Haug, Guido Freckmann
发表日期
2013/7
期刊
Journal of Diabetes Science and Technology
卷号
7
期号
4
页码范围
833-841
简介
Background
This study aimed at evaluating and comparing the performance of a new generation of continuous glucose monitoring (CGM) system versus other CGM systems, under daily lifelike conditions.
Methods
A total of 10 subjects (7 female) were enrolled in this study. Each subject wore two Dexcom G4™ CGM systems in parallel for the sensor lifetime specified by the manufacturer (7 days) to allow assessment of sensor-to-sensor precision. Capillary blood glucose (BG) measurements were performed at least once per hour during daytime and once at night. Glucose excursions were induced on two occasions. Performance was assessed by calculating the mean absolute relative difference (MARD) between CGM readings and paired capillary BG readings and precision absolute relative difference (PARD), i.e., differences between paired CGM readings.
Results
Overall aggregate MARD was 11.0% (n = 2392 …