M Kidoh, T Nakaura, S Nakamura, S Tokuyasu, H Osakabe, K Harada, Y Yamashita
发表日期
2014/1/1
期刊
Clinical radiology
卷号
69
期号
1
页码范围
e11-e16
出版商
WB Saunders
简介
Aim
To evaluate the image quality of O-MAR (Metal Artifact Reduction for Orthopedic Implants) for dental metal artefact reduction.
Materials and methods
This prospective study received institutional review board approval and written informed consent was obtained. Thirty patients who had dental implants or dental fillings were included in this study. Computed tomography (CT) images were obtained through the oral cavity and neck during the portal venous phase. The system reconstructed the O-MAR-processed images in addition to the uncorrected images. CT attenuation and image noise of the soft tissue of the oral cavity were compared between the O-MAR and the uncorrected images. Qualitative analysis was undertaken between the two image groups.
Results
The image noise of the O-MAR images was significantly lower than that of the uncorrected images (p < 0.01). O-MAR offered plausible attenuations of soft …