作者
A Hagiwara, M Hori, K Yokoyama, MY Takemura, C Andica, T Tabata, K Kamagata, M Suzuki, KK Kumamaru, M Nakazawa, N Takano, H Kawasaki, N Hamasaki, A Kunimatsu, S Aoki
发表日期
2017/2/1
期刊
American journal of neuroradiology
卷号
38
期号
2
页码范围
257-263
出版商
American Journal of Neuroradiology
简介
BACKGROUND AND PURPOSE
Synthetic MR imaging enables the creation of various contrast-weighted images including double inversion recovery and phase-sensitive inversion recovery from a single MR imaging quantification scan. Here, we assessed whether synthetic MR imaging is suitable for detecting MS plaques.
MATERIALS AND METHODS
Quantitative and conventional MR imaging data on 12 patients with MS were retrospectively analyzed. Synthetic T2-weighted, FLAIR, double inversion recovery, and phase-sensitive inversion recovery images were produced after quantification of T1 and T2 values and proton density. Double inversion recovery images were optimized for each patient by adjusting the TI. The number of visible plaques was determined by a radiologist for a set of these 4 types of synthetic MR images and a set of conventional T1-weighted inversion recovery, T2-weighted, and FLAIR …