作者
Yue Zhao, Lina Qiu, Yunlong Sun, Chong Huang, Ting Li
发表日期
2017/11/1
期刊
Biomedical optics express
卷号
8
期号
11
页码范围
5151-5159
出版商
Optica Publishing Group
简介
Extinction coefficient (ε) is a critical parameter for quantification of oxy-, deoxy-, and total-hemoglobin concentrations (Δ[HbO_2], Δ[Hb], Δ[tHb]) from optical measurements of Near-infrared spectroscopy (NIRS). There are several different ε data sets which were frequently used in NIRS quantification. A previous study reported that even a small variation in ε could cause a significant difference in hemodynamic measurements. Apparently the selection of an optimal ε data set is important for NIRS. We conducted oxygen-state-varied and blood-concentration-varied model experiments with 57 human blood samples to mimic tissue hemodynamic variations. Seven reported ε data sets were evaluated by comparisons between quantifications and assumed values. We found that the Moaveni et al (1970)’ ε data set was the optimal one, the NIRS quantification varied significantly among different ε data sets and parameter Δ[tHb …
引用总数
20182019202020212022202320241334179103