作者
Yu Lin, Chong Huang, Daniel Irwin, Lian He, Yu Shang, Guoqiang Yu
发表日期
2014/3/24
期刊
Applied Physics Letters
卷号
104
期号
12
出版商
AIP Publishing
简介
This study extended our recently developed noncontact diffuse correlation spectroscopy flowmetry system into noncontact diffuse correlation tomography (ncDCT) for three-dimensional (3-D) flow imaging of deep tissue. A linear array of 15 photodetectors and two laser sources connected to a mobile lens-focusing system enabled automatic and noncontact scanning of flow in a region of interest. These boundary measurements were combined with a finite element framework for DCT image reconstruction implemented into an existing software package. This technique was tested in computer simulations and using a tissue-like phantom with anomaly flow contrast design. The cylindrical tube-shaped anomaly was clearly reconstructed in both simulation and phantom. Recovered and assigned flow contrast changes in anomaly were found to be highly correlated: regression slope= 1.00, R 2= 1.00, and p< 10− 5 in …
引用总数
201520162017201820192020202120222023202476310794861
学术搜索中的文章