作者
Sandra Franco, José B Almeida, M Parafita
发表日期
2000/9/1
来源
Journal of Refractive Surgery
卷号
16
期号
5
页码范围
S661-S663
出版商
SLACK Incorporated
简介
Corneal thickness measurements find applications in areas such as diagnosis and management of corneal disorders and corneal surgery. We present a technique for the measurement of corneal thickness using a CCD camera mounted on a slit-lamp biomicroscope and common image processing software. Thickness measurements were performed on three RGP contact lenses of known thickness with an average error of 5 µm, SD of 8 µm. Measurements of a living cornea yielded a thickness of 505 µm and a SD of 8 µm for 10 consecutive measurements; this is an acceptable value. [J Refract Surg 2000;16:S661-S663]
引用总数