简介
PURPOSE: Recently the authors presented a technique that allows measurement of the corneal thickness along a vertical meridian from optical sections obtained using a slit-lamp microscope. We present a new system that allows rotary scanning of the cornea and consequent measurement of the corneal thickness along any meridian.
METHOD: The corneal thickness along two perpendicular meridians was measured from optical sections obtained with a new rotary scanning system. To obtain the optical sections, the cornea is illuminated with a light beam that is previously expanded in a fan by a small cylindrical lens rotated on a plane-containing axis to explore the cornea. The light diffused by both corneal surfaces is collected by two video cameras placed at an angle with the light beam. The axes of the two cameras define, with the visual axis, two planes normal to each other. After image acquisition, a processing …