简介
Geant4 application for tomographic emission (GATE) is a generic Monte Carlo simulation platform based on a general-purpose code GEANT4 and designed to simulate positron emission tomography (PET) and single photon emission tomography systems. Monte Carlo simulations are used in nuclear medicine to model imaging systems and develop and assess tomographic reconstruction algorithms and correction methods for improved image quantification. The purpose of this study is to validate two GATE models of the commercial available PET scanner HR+ and the PET/CT Biograph 2. The geometry of the system components has been described in GATE, including detector ring, crystal blocks, PMTs etc. The energy and spatial resolution of the scanners as given by the manufacturers have been taken into account. The GATE simulated results are compared directly to experimental data obtained using a number …