作者
Joao D Dias, Elaine I Haney, Blesy A Mathew, Carlos G Lopez-Espina, Adrian W Orr, Mark A Popovsky
发表日期
2017/4/1
期刊
Archives of pathology & laboratory medicine
卷号
141
期号
4
页码范围
569-577
出版商
the College of American Pathologists
简介
Context
Thromboelastography (TEG) is a whole blood, real-time analyzer measuring the viscoelastic properties of the hemostasis process and allowing for individualized goal-directed therapy. However, routine use of TEG requires validation of sample storage effect on clot parameters.
Objectives
To establish the minimum time required for equilibration time and the maximum time for sample storage for all commercially available TEG tests for the new-generation TEG 6s and to determine how those times compare with the older generation TEG 5000.
Design
Citrated and heparinized whole blood samples obtained from 20 healthy donors were analyzed for clot parameters at multiple time points for both the TEG 6s and the TEG 5000. Samples were activated with the citrated multichannel cartridge or the platelet-mapping cartridge in the TEG 6s or with recalcified …