作者
Mayss Alreem Nizar Hammed Mustafa F. Mahmood,Huda Farooq
发表日期
2022
期刊
Indonesian Journal of Electrical Engineering and Computer Science
卷号
25
期号
2
页码范围
980-988
简介
The electroretinogram (ERG) is an electrophysiological recording method that measures the retinal electrical potential. The electrical reaction is quantified by electrical interaction of the indicator electrode with the cornea or at various levels inside the retina. However, such ERG systems suffer from certain limitations and challenges, such as high cost, low a/b-wave amplitude, and the outcomes do not provide any information about patients. In this work, we designed and implemented a real-time prototype for an ERG system for measuring eye waves via diode-transistor logic (DTL)-electrode and AD624AD-model. In addition, a graphical user interface (GUI) via virtual instrument engineering workbench (LabVIEW) was used. The developed system achieved high amplitude for ERG a/b-waves of about 100 and 700 mV. In terms of a/b-waves in the system, the findings show that this study has good results for optimizing the measurement of ERG signals. The method showed satisfactory accuracy of about 92.5% for 10 participants aged 20-60 years and comprising both genders.