作者
Éric Bavu, Jean-Luc Gennisson, Mathieu Couade, Jeremy Bercoff, Vincent Mallet, Mathias Fink, Anne Badel, AnaÏs Vallet-Pichard, Bertrand Nalpas, Mickaël Tanter, Stanislas Pol
发表日期
2011/9/1
期刊
Ultrasound in medicine & biology
卷号
37
期号
9
页码范围
1361-1373
出版商
Elsevier
简介
Supersonic shear imaging (SSI) has recently been demonstrated to be a repeatable and reproducible transient bidimensional elastography technique. We report a prospective clinical evaluation of the performances of SSI for liver fibrosis evaluation in 113 patients with hepatitis C virus (HCV) and a comparison with FibroScan (FS). Liver elasticity values using SSI and FS ranged from 4.50 kPa to 33.96 kPa and from 2.60 kPa to 46.50 kPa, respectively. Analysis of variance (ANOVA) shows a good agreement between fibrosis staging and elasticity assessment using SSI and FS (p < 10−5). The areas under receiver operating characteristic (ROC) curves for elasticity values assessed from SSI were 0.948, 0.962 and 0.968 for patients with predicted fibrosis levels F ≥ 2, F ≥ 3 and F = 4, respectively. These values are compared with FS area under the receiver operating characteristic curve (AUROC) of 0.846, 0.857 and 0 …
引用总数
2011201220132014201520162017201820192020202120222023202472372757859505132322621168
学术搜索中的文章