Performance of a new integrated computer-assisted system (CADe/CADx) for detection and characterization of colorectal neoplasia

J Weigt, A Repici, G Antonelli, A Afifi, L Kliegis… - …, 2022 - thieme-connect.com
J Weigt, A Repici, G Antonelli, A Afifi, L Kliegis, L Correale, C Hassan, H Neumann
Endoscopy, 2022thieme-connect.com
Background Use of artificial intelligence may increase detection of colorectal neoplasia at
colonoscopy by improving lesion recognition (CADe) and reduce pathology costs by
improving optical diagnosis (CADx). Methods A multicenter library of≥ 200 000 images from
1572 polyps was used to train a combined CADe/CADx system. System testing was
performed on two independent image sets (CADe: 446 with polyps, 234 without; CADx: 267)
from 234 polyps, which were also evaluated by six endoscopists (three experts, three non …
Background Use of artificial intelligence may increase detection of colorectal neoplasia at colonoscopy by improving lesion recognition (CADe) and reduce pathology costs by improving optical diagnosis (CADx).
Methods A multicenter library of ≥ 200 000 images from 1572 polyps was used to train a combined CADe/CADx system. System testing was performed on two independent image sets (CADe: 446 with polyps, 234 without; CADx: 267) from 234 polyps, which were also evaluated by six endoscopists (three experts, three non-experts).
Results CADe showed sensitivity, specificity, and accuracy of 92.9 %, 90.6 %, and 91.7 %, respectively. Experts showed significantly higher accuracy and specificity, and similar sensitivity, while non-experts + CADe showed comparable sensitivity but lower specificity and accuracy than CADe and experts. CADx showed sensitivity, specificity, and accuracy of 85.0 %, 79.4 %, and 83.6 %, respectively. Experts showed comparable performance, whereas non-experts + CADx showed comparable accuracy but lower specificity than CADx and experts.
Conclusions The high accuracy shown by CADe and CADx was similar to that of experts, supporting further evaluation in a clinical setting. When using CAD, non-experts achieved a similar performance to experts, with suboptimal specificity.