studied at the scale of the automated reaction. Among the buffers tested, HEPES 0.3 M pH 4
allowed both high radiochemical purity (RCP) and radiochemical yield (RCY), without
prepurification of 68 Ga but after final purification of [68 Ga] Ga-FAPI-46 on a C 18 cartridge.
A longer reaction time did not show significant benefit on the RCP, while higher loads of
FAPI-46 and gentisic acid as anti-radiolysis compound allowed better RCY.