A novel deep learning based automatic auscultatory method to measure blood pressure

F Pan, P He, F Chen, J Zhang, H Wang… - International journal of …, 2019 - Elsevier
International journal of medical informatics, 2019Elsevier
Background It is clinically important to develop innovative techniques that can accurately
measure blood pressures (BP) automatically. Objectives This study aimed to present and
evaluate a novel automatic BP measurement method based on deep learning method, and
to confirm the effects on measured BPs of the position and contact pressure of stethoscope.
Methods 30 healthy subjects were recruited. 9 BP measurements (from three different
stethoscope contact pressures and three repeats) were performed on each subject. The …
Background
It is clinically important to develop innovative techniques that can accurately measure blood pressures (BP) automatically.
Objectives
This study aimed to present and evaluate a novel automatic BP measurement method based on deep learning method, and to confirm the effects on measured BPs of the position and contact pressure of stethoscope.
Methods
30 healthy subjects were recruited. 9 BP measurements (from three different stethoscope contact pressures and three repeats) were performed on each subject. The convolutional neural network (CNN) was designed and trained to identify the Korotkoff sounds at a beat-by-beat level. Next, a mapping algorithm was developed to relate the identified Korotkoff beats to the corresponding cuff pressures for systolic and diastolic BP (SBP and DBP) determinations. Its performance was evaluated by investigating the effects of the position and contact pressure of stethoscope on measured BPs in comparison with reference manual auscultatory method.
Results
The overall measurement errors of the proposed method were 1.4 ± 2.4 mmHg for SBP and 3.3 ± 2.9 mmHg for DBP from all the measurements. In addition, the method demonstrated that there were small SBP differences between the 2 stethoscope positions, respectively at the 3 stethoscope contact pressures, and that DBP from the stethoscope under the cuff was significantly lower than that from outside the cuff by 2.0 mmHg (P < 0.01).
Conclusion
Our findings suggested that the deep learning based method was an effective technique to measure BP, and could be developed further to replace the current oscillometric based automatic blood pressure measurement method.