FRIDA will be a near infrared imager and integral field spectrograph covering the wavelength range from 0.9 to 2.5 microns. Primary observing modes are: direct imaging and integral field spectroscopy. This paper describes the main advances in the development of the electronics and control system for both the mechanisms and house-keeping of FRIDA. In order to perform several tests of mechanisms in both room and cryogenic environments, a set of programs had been developed. All variables of the vacuum control system were determined and the main control structure based on one Programmable Logic Controller (PLC) had been established. A key function of the FRIDA’s control system is keeping the integrity of cryostat during all processes, so we have designed a redundant heating control system which will be in charge of avoiding cryostat inner overheating. In addition, some improvements of cryogenic and room temperature cabling structure are described.