Design and Implementation Control of PID Controller of Dialysate Pump of Hemodialysis Machine

G Samuel, A Arifin, MH Fatoni… - … , Network, and Intelligent …, 2020 - ieeexplore.ieee.org
2020 International Conference on Computer Engineering, Network …, 2020ieeexplore.ieee.org
Peristaltic pump used in hemodialysis machine to transfer all liquid components needed to
make a dialysate. In general, dialysate made from water, acid and bicarbonate. By adjusting
a correct composition from those components, dialysate can be obtained. There are two
parameters that should considered, pH and electrical conductivity of dialysate. The aim of
this research is to design a PID controller for peristaltic pump in order to create an
appropriate dialysate. As a result, peristaltic pump that has 55 mL/min flowrate can make an …
Peristaltic pump used in hemodialysis machine to transfer all liquid components needed to make a dialysate. In general, dialysate made from water, acid and bicarbonate. By adjusting a correct composition from those components, dialysate can be obtained. There are two parameters that should considered, pH and electrical conductivity of dialysate. The aim of this research is to design a PID controller for peristaltic pump in order to create an appropriate dialysate. As a result, peristaltic pump that has 55 mL/min flowrate can make an appropriate dialysate in 6.36 minutes. When system got interference, the system can control the pump using feedback from sensor. PID controller succesfully control peristaltic pumps used in this system. In the future, this research can be integrated with other sub-system to make a hemodialysis machine.