Detection efficiency of NaI (Tl) detector in 511–1332 keV energy range

I Akkurt, K Gunoglu, SS Arda - Science and Technology of …, 2014 - Wiley Online Library
I Akkurt, K Gunoglu, SS Arda
Science and Technology of Nuclear Installations, 2014Wiley Online Library
As it is important to obtain accurate analytical result in an experimental research, this
required quality control of the experimental system. Gamma spectrometry system can be
used in a variety of different fields such as radiation and medical physics. In this paper the
absolute efficiency, peak to valley ratio, and energy resolution of a 3′′× 3′′ NaI (Tl)
detector were determined experimentally for 511, 662, 835, 1173, 1275, and 1332 keV
photon energies obtained from 22Na, 54Mn, 60Co, and 137Cs radioactive sources.
As it is important to obtain accurate analytical result in an experimental research, this required quality control of the experimental system. Gamma spectrometry system can be used in a variety of different fields such as radiation and medical physics. In this paper the absolute efficiency, peak to valley ratio, and energy resolution of a 3 × 3 NaI(Tl) detector were determined experimentally for 511, 662, 835, 1173, 1275, and 1332 keV photon energies obtained from 22Na, 54Mn, 60Co, and 137Cs radioactive sources.
Wiley Online Library
以上显示的是最相近的搜索结果。 查看全部搜索结果