Duration of Systole and Diastole for Hydrodynamic Testing of Prosthetic Heart Valves: Comparison Between ISO 5840 Standards and in vivo Studies

O Bazan, JP Ortiz - Brazilian journal of cardiovascular surgery, 2016 - SciELO Brasil
Brazilian journal of cardiovascular surgery, 2016SciELO Brasil
Objective: To complement the ISO 5840 standards concerning the duration of left ventricular
systole and diastole as a function of changes in heart rates according to in vivo studies from
the physiologic literature review. Methods: The systolic and diastolic durations from three in
vivo studies were compared with the durations of systole proposed by the ISO 5840: 2010
and ISO 5840-2: 2015 for hydrodynamic performance assessment of prosthetic heart valves.
Results: Based on the in vivo studies analyzed, the systolic durations proposed by the ISO …
Objective
To complement the ISO 5840 standards concerning the duration of left ventricular systole and diastole as a function of changes in heart rates according to in vivo studies from the physiologic literature review.
Methods
The systolic and diastolic durations from three in vivo studies were compared with the durations of systole proposed by the ISO 5840:2010 and ISO 5840-2:2015 for hydrodynamic performance assessment of prosthetic heart valves.
Results
Based on the in vivo studies analyzed, the systolic durations proposed by the ISO 5840 standard seemed consistent for 45 and 120 beats per minute (bpm), and showed diverse results for the 70 bpm condition.
Conclusion
Information on the realistic validation of the operation of left ventricular models for different heart rates were obtained.