Femoral stem incorporating a diamond cubic lattice structure: Design, manufacture and testing

B Jetté, V Brailovski, M Dumas, C Simoneau… - Journal of the …, 2018 - Elsevier
Journal of the Mechanical Behavior of Biomedical Materials, 2018Elsevier
Background The current total hip prostheses with dense femoral stems are considerably
stiffer than the host bones, which leads to such long-term complications as aseptic
loosening, and eventually, the need for a revision. Consequently, the lifetime of the
implantation does not match the lifetime expectation of young patients. Method A femoral
stem design featuring a porous structure is proposed to lower its stiffness and allow bone
tissue ingrowth. The porous structure is based on a diamond cubic lattice in which the pore …
Background
The current total hip prostheses with dense femoral stems are considerably stiffer than the host bones, which leads to such long-term complications as aseptic loosening, and eventually, the need for a revision. Consequently, the lifetime of the implantation does not match the lifetime expectation of young patients.
Method
A femoral stem design featuring a porous structure is proposed to lower its stiffness and allow bone tissue ingrowth. The porous structure is based on a diamond cubic lattice in which the pore size and the strut thickness are selected to meet the biomechanical requirements of the strength and the bone ingrowth. A porous stem and its fully dense counterpart are produced by laser powder-bed fusion using Ti-6Al-4V alloy. To evaluate the stiffness reduction, static testing based on the ISO standard 7206-4 is performed. The experimental results recorded by digital image correlation are analyzed and compared to the numerical model.
Results & conclusions
The numerical and experimental force-displacement characteristics of the porous stem show a 31% lower stiffness as compared to that of its dense counterpart. Moreover, the correlation analysis of the total displacement and equivalent strain fields allows the preliminary validation of the numerical model of the porous stem. Finally, the analysis of the surface-to-volume and the strength-to-stiffness ratios of diamond lattice structures allow the assessment of their potential as biomimetic constructs for load-bearing orthopaedic implants.
Elsevier
以上显示的是最相近的搜索结果。 查看全部搜索结果