Influence and reliability of lower-limb arterial occlusion pressure at different body positions

L Hughes, O Jeffries, M Waldron, B Rosenblatt… - PeerJ, 2018 - peerj.com
PeerJ, 2018peerj.com
Background Total arterial occlusive pressure (AOP) is used to prescribe pressures for
surgery, blood flow restriction exercise (BFRE) and ischemic preconditioning (IPC). AOP is
often measured in a supine position; however, the influence of body position on AOP
measurement is unknown and may influence level of occlusion in different positions during
BFR and IPC. The aim of this study was therefore to investigate the influence of body
position on AOP. Methods Fifty healthy individuals (age= 29±6 y) underwent AOP …
Background
Total arterial occlusive pressure (AOP) is used to prescribe pressures for surgery, blood flow restriction exercise (BFRE) and ischemic preconditioning (IPC). AOP is often measured in a supine position; however, the influence of body position on AOP measurement is unknown and may influence level of occlusion in different positions during BFR and IPC. The aim of this study was therefore to investigate the influence of body position on AOP.
Methods
Fifty healthy individuals (age = 29 ± 6 y) underwent AOP measurements on the dominant lower-limb in supine, seated and standing positions in a randomised order. AOP was measured automatically using the Delfi Personalised Tourniquet System device, with each measurement separated by 5 min of rest.
Results
Arterial occlusive pressure was significantly lower in the supine position compared to the seated position (187.00 ± 32.5 vs 204.00 ± 28.5 mmHg, p < 0.001) and standing position (187.00 ± 32.5 vs 241.50 ± 49.3 mmHg, p < 0.001). AOP was significantly higher in the standing position compared to the seated position (241.50 ± 49.3 vs 204.00 ± 28.5 mmHg, p < 0.001).
Discussion
Arterial occlusive pressure measurement is body position dependent, thus for accurate prescription of occlusion pressure during surgery, BFR and IPC, AOP should be measured in the position intended for subsequent application of occlusion.