On-line glucose monitoring by using microdialysis sampling and amperometric detection based on 'wired'glucose oxidase in carbon paste

E Csöregi, T Laurell, I Katakis, A Heller, L Gorton - Microchimica Acta, 1995 - Springer
Microchimica Acta, 1995Springer
In-vitro on-line glucose monitoring is described, based on microdialysis sampling and
amperometric detection operated in a flow-injection system. Samples were injected into a
two-electrode microcell containing an Ag/AgCl quasi-reference electrode and a glucose
enzyme electrode as the working electrode, operated at+ 0.15 V vs. Ag/AgCl. The enzyme
electrode is constructed by mixing the 'wired'glucose oxidase into carbon paste.{Poly [1-
vinylimidazole osmium (4, 4′-dimethylbipyridine) 2 Cl)]}+/2+ was used to 'wire'the enzyme …
Abstract
In-vitro on-line glucose monitoring is described, based on microdialysis sampling and amperometric detection operated in a flow-injection system. Samples were injected into a two-electrode microcell containing an Ag/AgCl quasi-reference electrode and a glucose enzyme electrode as the working electrode, operated at + 0.15 Vvs. Ag/AgCl. The enzyme electrode is constructed by mixing the ‘wired’ glucose oxidase into carbon paste. {Poly[1-vinylimidazole osmium(4,4′-dimethylbipyridine)2Cl)]}+/2+ was used to ‘wire’ the enzyme. The non-coated electrodes, cross-linked with poly(ethylene glycol) diglycidyl ether, responded linearly to glucose concentrations up to 60 mM, and were characterized by a sensitivity of 0.23 μA mM−1 cm−2, when operated in flow injection mode and of 5.4 μAmM –1 cm–2 in steady-state conditions. This sensitivity of the resulting enzyme electrode was 50% lower than that of similarly prepared but non-cross-linked electrodes. However, the cross-linked electrodes showed superior operational and storage stabilities, which were further improved by coating the electrodes with a negatively charged Eastman AQ film. An in-house designed microdialysis probe, equipped with a polysulphone cylindrical dialysis membrane, yielded a relative recovery of 50–60% at a perfusion rate of 2.5 μl/min–1 in a well stirred glucose solution. The on-line set up effectively rejected common interferences such as ascorbic acid and 4-acetaminophen when present at their physiological concentrations.
Springer
以上显示的是最相近的搜索结果。 查看全部搜索结果