radiometer (HACR), we have realized a scale of absolute spectral response between 406
and 920 nm. The HACR, an electrical-substitution radiometer operating at cryogenic
temperatures, achieves a combined relative standard uncertainty of 0.021%. Silicon
photodiode light-trapping detectors were calibrated against the HACR with a typical relative