events based on fitting experimental 2-D charge-profiles to two analytical models. First
results are presented regarding a 10× 10 LuYAP: Ce array of 2× 2× 10 mm 3 crystal pixels
coupled to a H10966 Hamamatsu 8× 8 multi-anode assembly under radio-isotopic
irradiations and from self-activity. Results show that a photo multiplier tube with cross plate
anode configuration would be preferable than a multi anode one due to uniformity, cost, and …