developed. The main improvements are:(1) precise control of temperature using Peltier
modules,(2) a well sealed cell, which can prevent evaporation of immersion oil, and (3)
automatic measurement using an image memory board. With this system, refractive index of
each single grain of glass or mineral can be feasibly determined with an accuracy as good
as±2× 10− 4.