Dosimetric IMRT verification with a flat‐panel EPID

B Warkentin, S Steciw, S Rathee, BG Fallone - Medical physics, 2003 - Wiley Online Library
B Warkentin, S Steciw, S Rathee, BG Fallone
Medical physics, 2003Wiley Online Library
A convolution‐based calibration procedure has been developed to use an amorphous
silicon flat‐panel electronic portal imaging device (EPID) for accurate dosimetric verification
of intensity‐modulated radiotherapy (IMRT) treatments. Raw EPID images were
deconvolved to accurate, high‐resolution 2‐D distributions of primary fluence using a scatter
kernel composed of two elements: a Monte Carlo generated kernel describing dose
deposition in the EPID phosphor, and an empirically derived kernel describing optical …
A convolution‐based calibration procedure has been developed to use an amorphous silicon flat‐panel electronic portal imaging device (EPID) for accurate dosimetric verification of intensity‐modulated radiotherapy (IMRT) treatments. Raw EPID images were deconvolved to accurate, high‐resolution 2‐D distributions of primary fluence using a scatter kernel composed of two elements: a Monte Carlo generated kernel describing dose deposition in the EPID phosphor, and an empirically derived kernel describing optical photon spreading. Relative fluence profiles measured with the EPID are in very good agreement with those measured with a diamond detector, and exhibit excellent spatial resolution required for IMRT verification. For dosimetric verification, the EPID‐measured primary fluences are convolved with a Monte Carlo kernel describing dose deposition in a solid water phantom, and cross‐calibrated with ion chamber measurements. Dose distributions measured using the EPID agree to within 2.1% with those measured with film for open fields of and Predictions of the EPID phantom scattering factors based on our scatter kernels are within 1% of the measured for open field sizes of up to Pretreatment verifications of step‐and‐shoot IMRT treatments using the EPID are in good agreement with those performed with film, with a mean percent difference of for three IMRT treatments (24 fields).
Wiley Online Library
以上显示的是最相近的搜索结果。 查看全部搜索结果

Google学术搜索按钮

example.edu/paper.pdf
搜索
获取 PDF 文件
引用
References