Fluorous-phase ion-selective pH electrodes: electrode body and ionophore optimization for measurements in the physiological pH range

XV Chen, MPS Mousavi, P Bühlmann - ACS omega, 2020 - ACS Publications
ACS omega, 2020ACS Publications
Because of their low polarity and polarizability, fluorous sensing membranes are both
hydrophobic and lipophobic and exhibit very high ion selectivities. Here, we report on a new
fluorous-membrane ion-selective electrode (ISE) with a wide sensing range centered
around physiologically relevant pH values. The fluorophilic tris [perfluoro (octyl) butyl] amine
(N [(CH2) 4Rf8] 3) was synthesized and tested as a new H+ ionophore using a redesigned
electrode body that provides excellent mechanical sealing and much improved …
Because of their low polarity and polarizability, fluorous sensing membranes are both hydrophobic and lipophobic and exhibit very high ion selectivities. Here, we report on a new fluorous-membrane ion-selective electrode (ISE) with a wide sensing range centered around physiologically relevant pH values. The fluorophilic tris[perfluoro(octyl)butyl]amine (N[(CH2)4Rf8]3) was synthesized and tested as a new H+ ionophore using a redesigned electrode body that provides excellent mechanical sealing and much improved measurement reliability. In a challenging 1 M KCl background, these fluorous-phase ISEs exhibit a sensing range from pH 2.2 to 11.2, which is one of the widest working ranges reported to date for ionophore-based H+ ISEs. High selectivities against common interfering ions such as K+, Na+, and Ca2+ were determined (selectivity coefficients: logKH, Kpot = – 11.6; logKH, Napot = – 12.4; logKH, Capot < – 10.2). The use of the N[(CH2)4Rf8]3 ionophore with its −(CH2)4– spacers separating the amino group from the strongly electron-withdrawing perfluorooctyl groups improved the potentiometric selectivity as compared to the less basic tris[perfluoro(octyl)propyl]amine ionophore. The use of N[(CH2)4Rf8]3 also made the ISE less prone to counter anion failure (i.e., Donnan failure) at low pH than the use of tris[perfluoro(octyl)pentyl]amine with its longer −(CH2)5– spacers, which more effectively shield the amino center from the perfluorooctyl groups. In addition, we exposed both conventional plasticized PVC-phase pH ISEs and fluorous-phase pH ISEs to 10% serum for 5 days. Results show that the PVC-phase ISEs lost selectivity while their fluorous-phase counterparts did not.
ACS Publications
以上显示的是最相近的搜索结果。 查看全部搜索结果