Massage that pocket: a case of inappropriate shocks after subcutaneous ICD placement

A Kanwal, J Kaplan, A Saini - Journal of the American College of …, 2020 - jacc.org
A Kanwal, J Kaplan, A Saini
Journal of the American College of Cardiology, 2020jacc.org
… Air entrapment can lead to inappropriate shocks in S-ICD patients. Adequate pocket
massage before closure is critical step during implant. CXR may be critical at diagnoses in
patients with inappropriate shocks. The condition is self-resolving, and the device can be
temporarily reprogrammed to avoid …
Background
Subcutaneous ICD (S-ICD) has recently emerged as an alternative to transvenous ICD. However, inappropriate ICD therapy remains a concern.