Simultaneous measurement of hepatic and splenic stiffness using MR elastography: preliminary experience

HA Dyvorne, GH Jajamovich, C Besa, N Cooper… - Abdominal …, 2015 - Springer
Abdominal imaging, 2015Springer
Purpose To compare MR elastography (MRE) using a single and a dual driver excitation for
the quantification of hepatic and splenic stiffness (HS and SS), and to investigate the
performance of HS and SS measured with single or dual driver excitation for the detection of
liver cirrhosis in subjects with liver disease. Patients and methods This prospective HIPAA
compliant and IRB approved study involved 49 subjects who underwent MRE at 3.0 T,
comparing three different acquisition methods (single driver on the liver, single driver on the …
Purpose
To compare MR elastography (MRE) using a single and a dual driver excitation for the quantification of hepatic and splenic stiffness (HS and SS), and to investigate the performance of HS and SS measured with single or dual driver excitation for the detection of liver cirrhosis in subjects with liver disease.
Patients and methods
This prospective HIPAA compliant and IRB approved study involved 49 subjects who underwent MRE at 3.0T, comparing three different acquisition methods (single driver on the liver, single driver on the spleen and dual driver acoustic excitation). A Mann–Whitney test was used to assess changes in stiffness values. Bland–Altman analysis was used to compare single and dual driver configurations for each organ. Performance for detection of liver cirrhosis was assessed using ROC analysis. Pearson correlation was used to estimate the dependence of HS and SS on spleen size.
Results
There were 40 noncirrhotic and 9 cirrhotic patients. There was good agreement between stiffness values measured with a single or a dual driver (Bland–Altman limits of agreement −14.3 % to 18.9 % and −18.1 % to 29.7 %, CV 6.4 % and 9.4 %, for HS and SS. respectively). HS and SS were higher in subjects with liver cirrhosis (p < 0.001), with excellent detection performance (AUROC range 0.87–0.93). SS correlated strongly with spleen size (r = 0.69, p < 0.001), while HS showed weak correlation (r = 0.38, p = 0.006).
Conclusion
Using a dual acoustic driver configuration, hepatic and splenic stiffness can be simultaneously estimated with good concordance with single driver measurement.